Kikusui PLZ150U DC Electronic Load 1.5-150V 0-30A

£920.00 (£1,104.00 inc VAT)

3 in stock (can be backordered)

Availability: Stock models delivered next working day (UK only) for orders placed before 16:30 (Mon - Fri)

Description


Kikusui PLZ150UA DC Electronic Load

The PLZ150UA (1.5-150V 0-30A) Electronic Load is a compact, high-performance, multi-channel electronic load unit that has three operation modes: CC (constant current), CR (constant resistance), and CV (constant voltage). The PLZ-U consists of a frame and load units. Load units are inserted into the frame (control section) for operation. The PLZ-30F and PLZ-50F Frames can accommodate up to 3 channels and 5 channels of load units, respectively. The current capacity or power capacity can be increased through a parallel operation of load units. The load units can be combined to produce a power capacity from 75W to 750W (when five PLZ150Us are installed into the PLZ-50F). In addition to high-speed response and highly precise pre-set resolution, various functions are provided including the soft start function, variable slew rates, switching function, ABC pre-set memories, four set-up memories, and sequence function. GPIB and RS232C as fitted as standard.

Features:

  • Slew rate of 2.4 A/?s in the rising and falling edges in CC mode (PLZ150U)
  • Built-in three ranges; voltmeter, ammeter, and wattmeter functions that provide readings of up to five digits
  • The current slew rate can be changed continuously in constant current and constant resistance modes.
  • Supports 0-V input – an indispensable feature for testing single cell fuel cells.
  • Individual units (channels) can operate either independently or in synchronisation.
  • Up to five load units of the same model can be operated in parallel.
  • Up to three values can be stored in memory for each most frequently used operation mode and range.
  • Equipped with various types of protection circuits (over voltage protection, over current protection, over power protection, over heat protection, under voltage protection, and reverse connection protection).
  • Supports the GPIB and RS-232C interfaces as standard.
  • External control is available to turn on or off the output.

Documentation

PLZ-U/UA Specifications
ModelPLZ150UPLZ70UA
Ratings
Operating Voltage (DC)1.5 V to 150 V0 V to 150 V
Current/Power
H30 A/150 W15 A/75 W
M3 A/150 W1.5 A/75 W
L300 mA/45 W150 mA/22.5 W
Isolation voltage of the load input terminal500VDC500VDC
Withstand voltage between load input terminal channels500VDC500VDC
Minimum start voltage0.3 V or greater0.3 V or greater
CC Mode
Operating Range
H0 A to 30 A0 A to 15 A
M0 A to 3 A0 A to 1.5 A
L0 A to 300 mA0 A to 150 mA
Selectable Range0 % to 105 % of f.s0 % to 105 % of f.s
Resolution
H2mA1mA
M0.2mA0.1mA
L0.02mA0.01mA
Accuracy of setting (H, M and L)±(0.2 % of set + 0.2 % of f.s) + Vin / 500 kΩ±(0.2 % of set + 0.2 % of f.s) + Vin / 500 kΩ
Input voltage variation
H2mA2mA
M1mA1mA
L0.1mA0.1mA
Ripplerms: 3mA
p-p: 30mA
rms: 3mA
p-p: 30mA
CR Mode
Operating range
The value inside parentheses is
the conductance.
HOPEN to 50 mΩ (0 S to 20 S)OPEN to 100 mΩ (0 S to 10 S)
MOPEN to 500 mΩ (0 S to 2 S)OPEN to 1 Ω (0 S to 1 S)
LOPEN to 5 Ω (0 S to 200 mS)OPEN to 10 Ω (0 S to 100 mS)
Selectable Range0 % to 105 % of f.s0 % to 105 % of f.s
Resolution
H0.2 mS (0 S to 2 S)

2 mS (2 S to 20 S)

0.1 mS (0 S to 1 S)

1 mS (1 S to 10 S)

M20 μS (0 S to 200 mS)
0.2 mS (200 mS to 2 S)
10 μS (0 S to 100 mS)
0.1 mS (100 mS to 1 S)
L2 μS(0 S to 20 mS)
20 μS(20 mS to 200 mS)
1 μS(0 S to 10 mS)
10 μS(10 mS to 100 mS)
Accuracy of setting±(0.5 % of set + 0.5 % of f.s ) + Vin/500 kΩ±(0.5 % of set + 0.5 % of f.s ) + Vin/500 kΩ

Note* Specifications are liable to change. For the full list of Specifications, please see the products datasheet located under the ‘Documentation’ tab.

Reviews

There are no reviews yet.

Only logged in customers who have purchased this product may leave a review.