Rigol RSA3030 Real-Time Spectrum Analyser

The RSA3030 is a 9kHz to 3GHz Real-Time Spectrum Analyser. The Rigol RSA3000 series real-time spectrum analysers combines the power of a high-performance conventional spectrum analyser with superior real-time performance. Available in 3.0GHz and 4.5GHz models, all RSA3000 series analysers provide a resolution bandwidth (RBW) down to 1Hz (optional), a noise floor as low as -161dBm/Hz (typ.) and phase noise of -102dBc/Hz @ 10kHz offset (typ., f = 1GHz). In addition, the analyser features up to 40MHz of real-time bandwidth (optional), a minimum probability of intercept (POI) of 7.45µs, 7-rich visualisation modes and powerful triggering capability. The Rigol RSA3000 series is a complete analysis package for engineers developing and integrating todays wireless technologies utilising complex and dynamic modulation schemes.

Features:

  • Displayed average noise level (DANL): <-161dBm (typical) (With Optional Pre-Amp On)
  • Phase noise: <-102dBc/Hz @ 10kHz (typical, f = 1GHz)
  • Level measurement uncertainty: <1.0dB
  • Built-in tracking generator (‘-TG’ models only)
  • Minimum RBW: 1Hz (optional)
  • EMC filter and quasi-peak detector
  • Various measurement functions
  • Multiple measurement modes
  • Up to 40MHz real-time analysis bandwidth
  • Multiple trigger modes and trigger masks
  • Density, spectrogram, and other display modes
  • 10.1” capacitive multi-touch screen, supporting touch gestures
  • USB, LAN, HDMI and other communication and display interfaces
  • Supplied with… Power cable & quick guide
  • Options include… Preamplifier, high-stability reference, 25MHz real-time / analysis bandwidth, 40MHz real-time / analysis bandwidth, advanced measurement kit.
  • Optional PC software… Ultra Spectrum, S1210 EMI Pre-Compliance Test
  • Accessories include… Rack mount kit

Rigol RSA3000 Real-Time Spectrum Analyser Comparison Table

ModelFrequency RangeReal-Time BandwidthMinimum Resolution Bandwidth (RBW)Maximum CW RF PowerSSB Phase Noise (@10kHz Offset)DANL, PreAmplifier Off (TYP. 20MHz – 2.7GHz)DANL, PreAmplifier On (TYP. 20MHz – 2.7GHz)EMI Filter & Quasi-Peak Detector
RSA30309kHz – 3GHz10MHz (standard)
25MHz & 40MHz (optional)
10Hz (standard)
1Hz (optional)
30dBm (1W)<-102dBc/Hz typ.<-141dBm/Hz typ.<-161dBm/Hz typ.Optional
RSA3030-TG9kHz – 3GHz10MHz (standard)
25MHz & 40MHz (optional)
10Hz (standard)
1Hz (optional)
30dBm (1W)<-102dBc/Hz typ.<-141dBm/Hz typ.<-161dBm/Hz typ.Optional
RSA30459kHz – 4.5GHz10MHz (standard)
25MHz & 40MHz (optional)
10Hz (standard)
1Hz (optional)
30dBm (1W)<-102dBc/Hz typ.<-141dBm/Hz typ.<-161dBm/Hz typ.Optional
RSA3045-TG9kHz – 4.5GHz10MHz (standard)
25MHz & 40MHz (optional)
10Hz (standard)
1Hz (optional)
30dBm (1W)<-102dBc/Hz typ.<-141dBm/Hz typ.<-161dBm/Hz typ.Optional

Product Videos

Documentation

Measurement Mode
Frequency RangeRSA3015N: 9 kHz to 1.5 GHz
RSA3030/-TG/N: 9 kHz to 3.0 GHz
RSA3045/-TG/N: 9 kHz to 4.5 GHz
Reference Frequency10 MHz
Accuracy±[(time since last calibration × aging rate) + temperature stability + calibration
accuracy]
Initial Calibration
Accuracy
Standard <1 ppm
Option OCXO-C08 <0.1 ppm
Temperature
Stability
0℃ to 50℃ , with the reference 25℃
Standard <0.5 ppm
Option OCXO-C08 <0.005 ppm
Aging RateStandard <1 ppm/year
Option OCXO-C08 <0.03 ppm/year
Frequency
Marker Frequency Resolutionspan/(number of sweep points – 1)
Marker Frequency Uncertainty±(marker frequency readout × reference frequency accuracy + 1% × span + 10% ×
resolution bandwidth + marker frequency resolution)
Resolution1 Hz
Uncertainty±(marker frequency readout × reference frequency accuracy + counter resolution)
RangeStandard 0 Hz, 100 Hz to maximum frequency
Option RSA3000-BW1 0 Hz, 10 Hz to maximum frequency
Resolution2 Hz
Uncertainty±span/(number of sweep points – 1)
Carrier Offset1 kHz <-90 dBc/Hz (typical)
10 kHz <-100 dBc/Hz, <-102 dBc/Hz (typical)
100 kHz <-100 dBc/Hz, <-102 dBc/Hz (typical)
1 MHz <-110 dBc/Hz, <-112 dBc/Hz (typical)
Amplitude
Rangefc ≥ 10 MHz
DANL to +30 dBm
DC Voltage50 V
CW RF Power+30 dBm, attenuation ≥ 40 dB, preamp off.
-10 dBm, attenuation = 20 dB, preamp on.
CW RF Power+33 dBm (2 W)
Logarithmic Scale1 dB to 200 dB
Linear Scale0 to reference level
Number of Display Points801
Number of Traces6
Trace Detectornormal, pos-peak, neg-peak, sample, RMS average, voltage average, and
quasi-peak (Option RSA3000-EMC)
Trace Functionclear write, max hold, min hold, average, view, blank
Scale UnitdBm, dBmV, dBμV, nV, μV, mV, V, nW, μW, mW, W
General Specifications
Display
Typecapacitive multi-touch screen
Resolution1024 × 600 pixels
Size10.1”
Colour24-bit colour
Power
Input Voltage Range, AC100 V to 240 V (nominal)
AC Frequency45 Hz to 440 Hz
Power Consumption55 W (typical), max. 90 W with all options
Environment
TemperatureOperating Temperature
Range 0℃ to 50℃
Storage Temperature
Range -20℃ to 70℃
Humidity0℃ to 30℃ ≤ 95% RH
30℃ to 40℃ ≤ 75% RH
AltitudeOperating Height: below 3,048 m (10,000 feet)

Note*

Specifications are liable to change. For the full list of Specifications, please see the products datasheet located under the ‘Documentation’ tab.