Rental Kikusui TOS9303 AC/DC HiPot Tester with Insulation Resistance and Ground Bond Test

£295.00 (£354.00 inc VAT)

Prices Shown are Rental Price Per Week

1 in stock

Availability: Stock models delivered next working day (UK only) for orders placed before 16:30 (Mon - Fri)

Description


Rental Kikusui TOS9303 Safety Tester

The TOS9300 Series is a high performance electrical safety analyser that complies to a wide range of universal standards. HiPot, Insulation Resistance, Ground Bond, Leakage Current (touch current and protective conductor current) and partial discharge can all be tested. A total of 6 models are available for standard compliance tests in a wide variety of applications including R&D, quality assurance manufacturing lines and laboratory tests. The TOS9300 Series is a high performance electrical safety analyser that complies to a wide range of universal standards. HiPot, Insulation Resistance, Ground Bond, Leakage Current (touch current and protective conductor current) and partial discharge can all be tested. A total of 6 models are available for standard compliance tests in a wide variety of applications including R&D, quality assurance manufacturing lines and laboratory tests.

Features:

  • User-friendly 10Key configuration
  • Easy firmware updates via USB
  • LAN/USB/RS232C Standard Digital Interface
  • I/V Monitor Terminal (Analogue Monitor)
  • STATUS OUT Connector
  • SIGNAL I/O Connector
  • High Precision/High Resolution/High Speed
  • Programmable detection response speed
  • 7.2kV/100W DC HiPot Test
  • Positive Electrode/Negative Electrode Insulation Resistance Testing
  • Electric Discharge Function
  • AC/DC Earth Continuity testing up to 40A
  • EARTH FAULT Protection
  • Offset Cancel
  • Basic memory function
  • Calibration deadline notification
  • High-voltage scanner (option)

Specification:

  • Max output-voltage of AC HiPot testing: 5kV / 100mA(500VA)
  • Max output-voltage of DC HiPot testing: 5kV / 20mA, 7.2kV.13.9mA(100W)
  • Measurement range insulation resistance testing: 0.001MΩ to 100.0GΩ (DC -25V to -1000V / DC +50V to +7200V)
  • Measurement range of ground bond testing: 0.001Ω to 0.600Ω (3.0A to 42.0A)
  • Rise-time control / fall-time control function provided
  • Output voltage and panel settings are controlled through LAN/USB/RS232C

 

Product Videos

Documentation

TOS9300 Specifications
ModelTOS9300TOS9301TOS9302TOS9303TOS9303LC
Output Function
AC Output Section
Output range0.050 kV to 5.000 kV0.050 kV to 5.000 kV0.050 kV to 5.000 kV0.050 kV to 5.000 kV0.050 kV to 5.000 kV
Resolution1 V1 V1 V1 V1 V
Setting accuracy±(1.2 % of setting + 20 V) (at no load)±(1.2 % of setting + 20 V) (at no load)±(1.2 % of setting + 20 V) (at no load)±(1.2 % of setting + 20 V) (at no load)±(1.2 % of setting + 20 V) (at no load)
Max. rated load500 VA(5 kV / 100 mA)500 VA(5 kV / 100 mA)500 VA(5 kV / 100 mA)500 VA(5 kV / 100 mA)500 VA(5 kV / 100 mA)
Max. rated current100 mA (when the output voltage is 0.2 kV or higher)100 mA (when the output voltage is 0.2 kV or higher)100 mA (when the output voltage is 0.2 kV or higher)100 mA (when the output voltage is 0.2 kV or higher)100 mA (when the output voltage is 0.2 kV or higher)
Transformer rating500 VA500 VA500 VA500 VA500 VA
Output voltage waveformSineSineSineSineSine
Distortion2 % or less. (when the output voltage is 0.5 kV or higher and no load or a pure resistive load is connected)2 % or less. (when the output voltage is 0.5 kV or higher and no load or a pure resistive load is connected)2 % or less. (when the output voltage is 0.5 kV or higher and no load or a pure resistive load is connected)2 % or less. (when the output voltage is 0.5 kV or higher and no load or a pure resistive load is connected)2 % or less. (when the output voltage is 0.5 kV or higher and no load or a pure resistive load is connected)
Crest factor√2 ± 3 % (800 V or more)√2 ± 3 % (800 V or more)√2 ± 3 % (800 V or more)√2 ± 3 % (800 V or more)√2 ± 3 % (800 V or more)
Frequency50 Hz / 60 Hz50 Hz / 60 Hz50 Hz / 60 Hz50 Hz / 60 Hz50 Hz / 60 Hz
Accuracy±0.1 %±0.1 %±0.1 %±0.1 %±0.1 %
Voltage regulation±3 % or less (when changing from maximum rated load to no load)±3 % or less (when changing from maximum rated load to no load)±3 % or less (when changing from maximum rated load to no load)±3 % or less (when changing from maximum rated load to no load)±3 % or less (when changing from maximum rated load to no load)
Short-circuit current200 mA or more (output voltage 0.5 kV or higher)200 mA or more (output voltage 0.5 kV or higher)200 mA or more (output voltage 0.5 kV or higher)200 mA or more (output voltage 0.5 kV or higher)200 mA or more (output voltage 0.5 kV or higher)
Output methodPWM switchingPWM switchingPWM switchingPWM switchingPWM switching
Start voltageThe voltage at the start of the test can be set.The voltage at the start of the test can be set.The voltage at the start of the test can be set.The voltage at the start of the test can be set.The voltage at the start of the test can be set.
Setting range0 % to 99 % of the test voltage0 % to 99 % of the test voltage0 % to 99 % of the test voltage0 % to 99 % of the test voltage0 % to 99 % of the test voltage
Resolution1 %1 %1 %1 %1 %
Output voltage monitor functionIf the output voltage exceeds ±(10 % of setting + 50 V), the output is turned off, and the protection function is activated.If the output voltage exceeds ±(10 % of setting + 50 V), the output is turned off, and the protection function is activated.If the output voltage exceeds ±(10 % of setting + 50 V), the output is turned off, and the protection function is activated.If the output voltage exceeds ±(10 % of setting + 50 V), the output is turned off, and the protection function is activated.If the output voltage exceeds ±(10 % of setting + 50 V), the output is turned off, and the protection function is activated.
DC Output Function
DC Output Section
Output voltage range0.050 kV to 7.200 kV0.050 kV to 7.200 kV0.050 kV to 7.200 kV0.050 kV to 7.200 kV0.050 kV to 7.200 kV
Resolution1 V1 V1 V1 V1 V
Setting accuracy±(1.2 % of setting + 20 V)±(1.2 % of setting + 20 V)±(1.2 % of setting + 20 V)±(1.2 % of setting + 20 V)±(1.2 % of setting + 20 V)
Max. rated load100 W (5 kV/20 mA, 7.2 kV/13.9 mA)100 W (5 kV/20 mA, 7.2 kV/13.9 mA)100 W (5 kV/20 mA, 7.2 kV/13.9 mA)100 W (5 kV/20 mA, 7.2 kV/13.9 mA)100 W (5 kV/20 mA, 7.2 kV/13.9 mA)
Max. rated current20 mA20 mA20 mA20 mA20 mA
Ripple
7.2 kV no load20 Vp-p (TYP)20 Vp-p (TYP)20 Vp-p (TYP)20 Vp-p (TYP)20 Vp-p (TYP)
Max. rated load50 Vp-p (TYP)50 Vp-p (TYP)50 Vp-p (TYP)50 Vp-p (TYP)50 Vp-p (TYP)
Voltage regulation1 % or less (when changing from maximum rated load to no load)1 % or less (when changing from maximum rated load to no load)1 % or less (when changing from maximum rated load to no load)1 % or less (when changing from maximum rated load to no load)1 % or less (when changing from maximum rated load to no load)
Short-circuit current100 mA (TYP) (200 mA peak)100 mA (TYP) (200 mA peak)100 mA (TYP) (200 mA peak)100 mA (TYP) (200 mA peak)100 mA (TYP) (200 mA peak)
Discharge functionForced discharge after test completion (discharge resistance: 125 kΩ)Forced discharge after test completion (discharge resistance: 125 kΩ)Forced discharge after test completion (discharge resistance: 125 kΩ)Forced discharge after test completion (discharge resistance: 125 kΩ)Forced discharge after test completion (discharge resistance: 125 kΩ)
Start voltageThe voltage at the start of the test can be set.The voltage at the start of the test can be set.The voltage at the start of the test can be set.The voltage at the start of the test can be set.The voltage at the start of the test can be set.
Setting range0 % to 99 % of the test voltage0 % to 99 % of the test voltage0 % to 99 % of the test voltage0 % to 99 % of the test voltage0 % to 99 % of the test voltage
Resolution1 %1 %1 %1 %1 %
Output voltage monitor functionIf the output voltage exceeds ±(10 % of setting + 50 V), the output is turned off, and the protection function is activated.If the output voltage exceeds ±(10 % of setting + 50 V), the output is turned off, and the protection function is activated.If the output voltage exceeds ±(10 % of setting + 50 V), the output is turned off, and the protection function is activated.If the output voltage exceeds ±(10 % of setting + 50 V), the output is turned off, and the protection function is activated.If the output voltage exceeds ±(10 % of setting + 50 V), the output is turned off, and the protection function is activated.
Measurement Function
Voltmeter
Measurement range0.00 kV to 7.50 kV AC/DC0.00 kV to 7.50 kV AC/DC0.00 kV to 7.50 kV AC/DC0.00 kV to 7.50 kV AC/DC0.00 kV to 7.50 kV AC/DC
Resolution0.1 V0.1 V0.1 V0.1 V0.1 V
Accuracy±(1.2 % of reading + 5 V)±(1.2 % of reading + 5 V)±(1.2 % of reading + 5 V)±(1.2 % of reading + 5 V)±(1.2 % of reading + 5 V)
ResponseCan be switched between true rms and mean-value response rms conversion.
Peak-value response in a separate system
(the peak-value response is for measuring the dielectric breakdown voltage while rising)
Can be switched between true rms and mean-value response rms conversion.
Peak-value response in a separate system
(the peak-value response is for measuring the dielectric breakdown voltage while rising)
Can be switched between true rms and mean-value response rms conversion.
Peak-value response in a separate system
(the peak-value response is for measuring the dielectric breakdown voltage while rising)
Can be switched between true rms and mean-value response rms conversion.
Peak-value response in a separate system
(the peak-value response is for measuring the dielectric breakdown voltage while rising)
Can be switched between true rms and mean-value response rms conversion.
Peak-value response in a separate system
(the peak-value response is for measuring the dielectric breakdown voltage while rising)
Hold functionThe voltage measurement after a test is finished is held while the pass/fail judgment is displayedThe voltage measurement after a test is finished is held while the pass/fail judgment is displayedThe voltage measurement after a test is finished is held while the pass/fail judgment is displayedThe voltage measurement after a test is finished is held while the pass/fail judgment is displayedThe voltage measurement after a test is finished is held while the pass/fail judgment is displayed
Ammeter
Measurement rangeAC: 0.00 mA to 110 mA, DC: 0.00 mA to 22 mA (Current including the active component and reactive component)AC: 0.00 mA to 110 mA, DC: 0.00 mA to 22 mA (Current including the active component and reactive component)AC: 0.00 mA to 110 mA, DC: 0.00 mA to 22 mA (Current including the active component and reactive component)AC: 0.00 mA to 110 mA, DC: 0.00 mA to 22 mA (Current including the active component and reactive component)AC: 0.00 mA to 110 mA, DC: 0.00 mA to 22 mA (Current including the active component and reactive component)
Accuracy±(1 % of reading + 2 μA) (active component)±(1 % of reading + 2 μA) (active component)±(1 % of reading + 2 μA) (active component)±(1 % of reading + 2 μA) (active component)±(1 % of reading + 2 μA) (active component)
ResponseCan be switched between true rms and mean-value response rms conversion.Can be switched between true rms and mean-value response rms conversion.Can be switched between true rms and mean-value response rms conversion.Can be switched between true rms and mean-value response rms conversion.Can be switched between true rms and mean-value response rms conversion.
Hold functionThe current measurement after a test is finished is held while the pass judgment is displayedThe current measurement after a test is finished is held while the pass judgment is displayedThe current measurement after a test is finished is held while the pass judgment is displayedThe current measurement after a test is finished is held while the pass judgment is displayedThe current measurement after a test is finished is held while the pass judgment is displayed
Offset cancel functionCancels up to 10 mA of the current flowing through the insulation resistance and stray capacitance components across
output cables and the like (resistance component only for DC tests). OFF function available.
Cancels up to 10 mA of the current flowing through the insulation resistance and stray capacitance components across
output cables and the like (resistance component only for DC tests). OFF function available.
Cancels up to 10 mA of the current flowing through the insulation resistance and stray capacitance components across
output cables and the like (resistance component only for DC tests). OFF function available.
Cancels up to 10 mA of the current flowing through the insulation resistance and stray capacitance components across
output cables and the like (resistance component only for DC tests). OFF function available.
Cancels up to 10 mA of the current flowing through the insulation resistance and stray capacitance components across
output cables and the like (resistance component only for DC tests). OFF function available.
CalibrationActive component: Calibrated with the rms of a sine wave using a pure resistive load.
Reactive component: Not calibrated.
Active component: Calibrated with the rms of a sine wave using a pure resistive load.
Reactive component: Not calibrated.
Active component: Calibrated with the rms of a sine wave using a pure resistive load.
Reactive component: Not calibrated.
Active component: Calibrated with the rms of a sine wave using a pure resistive load.
Reactive component: Not calibrated.
Active component: Calibrated with the rms of a sine wave using a pure resistive load.
Reactive component: Not calibrated.

Note* Specifications are liable to change. For the full list of Specifications, please see the products datasheet located under the ‘Documentation’ tab.