Model |
TOS5200 |
AC Output Section |
|
Output Range |
0.05 kV to 5.00 kV |
Setting Accuracy |
± (2 % of set + 20 V) [at no load] |
Setting Range |
0.00 kV to 5.50 kV |
Resolution |
10 V STEP |
Max. rated output |
500 VA (5 kV/100 mA) |
Max. rated voltage |
5kV |
Max. rated current |
100 mA [output voltage 0.5 kV or higher] |
Transformer rating |
500 VA |
Output voltage waveform |
Sine |
Distortion |
If the output voltage is 0.5 kV or more: 3 % or less (when no load or a pure resistive load is connected) |
Crest factor |
Within √2 ± 3 % (output voltage 800 V or higher, at no load) |
Frequency |
50 Hz or 60 Hz |
Accuracy |
± 0.5 % (excluding during voltage rise time) |
Voltage regulation |
10 % or less (when changing from maximum rated load to no load) |
Input voltage variation |
±0.3 % (5 kV when no load is connected; power supply voltage: 90 V to 250 V) |
Short-circuit current |
200 mA or more (when the output voltage is 1.0 kV or greater) |
Output method |
PWM switching |
Start voltage |
The voltage at the start of withstanding voltage tests can be set to 50 % of the test voltage. |
Limit voltage |
The test voltage upper limit can be set . AC: 0.00 kV to 5.50 kV |
Output voltage monitor feature |
If output voltage exceeds the specified value + 350 V or is lower than the specified value – 350 V,
output is turned off, and protective features are activated. |
Voltmeter (Digital) |
|
Measurement range |
0.000 kV to 6.500 kV AC |
Display |
□.□□□ kV |
Accuracy |
V < 500 V: ± (1.5 % of reading + 20 V), V ≥ 500 V: ±1.5 % of reading |
Response |
True rms, Average value response/rms display switchable |
Hold feature |
After a test is finished, the measured voltage is retained until the PASS or FAIL judgment is cleared |
Ammeter (Digital) |
|
Measurement range |
0.00 mA to 110 mA |
Display |
i < 1 mA 1 mA ≤ i < 10 mA 10 mA ≤ i < 100 mA 100 mA ≤ i
□.□□□ mA □.□□□ mA □□.□□ mA □□□.□ mA |
Accuracy |
1.00 mA ≤ i: ± (1.5 % of reading), i < 1.00 mA: ± (1.5 % of reading + 30 μA) |
Response |
True rms/ Mean-value response rms display Can be switched |
Hold feature |
After a test is finished, the measured current value is retained until the PASS judgment is cleared. |
General Specification |
|
Display |
LCD: LED back custom indicators |
Environment |
|
Installation location |
Indoors, at a height of up to 2000 m |
Spec guaranteed range temperature/humidity |
5 °C to 35 °C (41 °F to 95 °F)/20 %rh to 80 %rh (no condensation) |
Operating range temperature/humidity |
0 °C to 40 °C (32 °F to 104 °F)/20 %rh to 80 %rh (no condensation) |
Storage range temperature/humidity |
-20 °C to 70 °C (-4 °F to 158 °F)/90 %rh or less (no condensation) |
Power supply |
|
Nominal voltage range (allowable voltage range) |
100 VAC to 240 VAC (90 VAC to 250 VAC) |
Power consumption |
100 VA or less, 800 VA max. |
Allowable frequency range |
47 Hz to 63 Hz |
Insulation resistance (between AC LINE and the chassis) |
30 MΩ or more (500 VDC) |
Withstanding voltage (between AC LINE and the chassis) |
1500 Vac, 1 minute |
Earth continuity |
25 Aac, 0.1 Ω or less |
Dimensions |
320 (12.6″) (330(12.99″)) W × 132(5.2″) (150(5.91″)) H × 350(13.78″) (420(16.54″)) D |
Weight |
Approx. 14 kg (30.9 lbs) |