Kikusui SD009-PCR-LE/WE Power Line Disturbance Immunity Testing Software
“Quick Immunity Sequencer” (model name: SD009-PCR-LE/WE) is an application software for immunity testing with the AC/DC power supply PCR-WE/WEA2 series system, based on the power line disturbance standard (IEC61000-4 Series) for the immunity testing of the EMC standard. Not only can it be used for compliance testing based on the latest standards or for some types of preliminary testing, but the software can be also employed for advance checking in development phases and for immunity margin tests, because it allows extended testing conditions to be set as needed.
Capable of performing compliance testing applied to the IEC61000-4-11, 4-13, 4-14, 4-17, 4-27, 4-28, 4-29, 4-34 or to perform preliminary tests before the standard testing.
List of conformance to the EMC standard tests
✔ :Conforming as standard ▲ :Partially non-conforming ─ :Function not available
Standard | Item | Single-Phase Conforming | Three-Phase Conforming |
---|---|---|---|
IEC61000-1-11 Voltage dips, short interruptions and voltage variations | Voltage Dips | ✔ | ✔ |
Short Interruptions | ✔ | ✔ | |
Voltage Variations | ✔ | ✔ | |
IEC61000-4-13 Harmonics and inter-harmonics | Flat curve | ✔ | ✔ |
Over swing | ✔ | ✔ | |
Frequency sweep | ✔ | ✔ | |
Odd harmonics the order of which is not a multiple of 3 | ✔ | ✔ | |
Odd harmonics the order of which is a multiple of 3 | ✔ | ✔ | |
Even harmonics | ✔ | ✔ | |
Interharmonics | ✔ | ✔ | |
Meister curve | ✔ | ✔ | |
IEC61000-4-14 Voltage fluctuation | Voltage fluctuation | ✔ | ✔ |
Interval | ✔ | ✔ | |
IEC61000-4-17 Ripple on DC input power port | Single-phase rectifier circuit | ✔ | ─ |
Three-phase rectifier circuit | ✔ | ─ | |
IEC61000-4-27 Unbalance in units | Unbalance | ─ | ▲*1 |
IEC61000-4-28 Variation of power frequency for equipment with 16 A/phase | Frequency variation | ✔ | ✔ |
IEC61000-4-29 Voltage dips, short interruptions and voltage variations on DC | Voltage dips | ✔ | ─ |
Short interruptions | ▲*3 | ─ | |
Voltage variations | ✔ | ─ | |
IEC61000-4-34 Voltage dips, short interruptions and voltage variations for equipment with input current more than 16 A/phase | Voltage dips | ▲*2 | ▲*2 |
Short interruptions | ▲*2 | ▲*2 | |
Voltage variations | ✔ | ✔ |
* Immunity testing for units with 16 A/phase except for those required by IEC61000-4-34
*1. Capability of rapid change with 1 µs to 5 µs is required for 110 %, 95.2 %, 93.5 %, 90 %, 87 %, 80 %, 74 %, 71 %, 66 %. Preliminary test is capable since the voltage response of the PCR-WE/WE2 is 55 µs in FAST mode.
*2. The device between the range of 16 A to 75 A requires to have the capability of rapid change with 1 μs to 5 μs. The device exceeding 75 A does not require to have the capability of rapid change with 1μs to 5 μs. (It is relaxed to 1 µs to 50 µs for the device exceeding 75 A.)
*3. Must support output impedance greater than 100 kΩ. The PCR-WE/WE2 output impedance is less than 100 kΩ and therefore designed for preliminary testing purposes.