How long will the DS2000A Capture a Signal

Posted on: December 16th, 2020 by Doug Lovell

To determine how long an oscilloscope will capture a signal when it is operated in single trigger mode is determined by taking the instrument’s memory depth and dividing it by the instrument’s sample rate.  The memory depth can be changed within the instrument’s acquire menu and then the sample rate can be changed by adjusting the horizontal scaling.  The table below lists the different memory depth and sample rate combinations that are possible on the DS2000A series and lists the length of time the instrument is able of capturing with each possible combination.  

Note that the length of time listed is in seconds and this is for one channel.

Memory Depth
Sample Rate1400014000014000001400000056000000
10001414014001400056000
2000770700700028000
50002.828280280011200
100001.4141401405600
200000.77707002800
500000.282.8282801120
1000000.141.414140560
2000000.070.777028
5000000.0280.282.828112
10000000.0140.141.41456
20000000.0070.070.7728
50000000.00280.0280.282.811.2
100000000.00140.0140.141.45.6
200000000.00070.0070.070.72.8
500000000.000280.00280.0280.281.12
1000000000.000140.00140.0140.140.56
2000000000.000070.00070.0070.070.28
5000000000.0000280.000280.00280.0280.112
10000000000.0000140.000140.00140.0140.056
20000000000.0000070.000070.00070.0070.028