Meatest M133Ci 3F Power & Energy Calibrator

. (£ inc VAT)

Please call for Delivery Time on this Model.

Availability: Stock models delivered next working day (UK only) for orders placed before 16:30 (Mon - Fri)

Description


Meatest M133Ci 3F Power & Energy Calibrator

M133C is electric AC/DC power and energy calibrator for calibration of power meters, power transducers and generally all kinds of power measuring devices. AC voltage, current, power and energy functions have calibrated phase shift in frequency range from 15 Hz to 1 kHz with resolution as good as 0.001 Hz. Simulated electric power can be set independently in each phase of both voltage and current signals. Built-in 5½ process multimeter for transducer calibration is standard part of every M133C calibrator.

M133C is more than just a sophisticated power source. Using built-in process multimeter, both transducer inputs and outputs can be handled by the M133C at the same time so you can calibrate it more easily, using no other calibration equipment. Floating current outputs can be connected directly with voltage outputs to allow for 3-wire power meter calibration. One calibrator, many applications – saving your time, space and costs.

Features:

  • 0.01° phase shift accuracy
  • 15 – 1000 Hz output
  • Built-in multimeter for transducer calibration
  • Floating current outputs for 3-wire power meters
  • Voltage from Current function

Product Videos

Documentation

DC voltage1 – 280 V
± 150 ppm / year
AC voltage1 – 600 V, 15 – 1000 Hz
± 120 ppm / year
DC current5 mA – 30 A (1500 A with 140-50 Current Coil)
± 175 ppm / year
AC current5 mA – 30 A, 15 – 1000 Hz (1500 A with 140-50 Current Coil)
± 175 ppm / year
DC Power & Energy5 mW – 8.4 kW (420 kW with 140-50 Current Coil)
± 380 ppm / year
AC Power & Energy5 mW – 18 kW, 15 – 1000 Hz (900 kW with 140-50 Current Coil)
0.01° phase shift accuracy, power factor -1 to +1
Harmonic, interharmonic, modulation/flicker, dip/swell distrortion functions
± 370 ppm / year

Process Multimeter

DC voltage0 – 12 V
± 100 ppm / year
DC current0 – 25 mA
± 100 ppm / year
Frequency1 Hz – 15 kHz
± 50 ppm / year

Optional Accessories …