| Test Functions |
| Insulation resistance test |
Yes |
Yes |
Yes |
Yes |
| AC withstand (flash) test |
Not supported |
Not supported |
Not supported |
Not supported |
| DC withstand (flash) test |
Not supported |
Not supported |
Not supported |
Not supported |
| Impulse, earth bond, leakage current or partial-discharge test |
Not supported |
Not supported |
Not supported |
Not supported |
| Output |
| Output voltage |
25 Vdc to 1000 Vdc |
25 Vdc to 1000 Vdc |
25 Vdc to 1000 Vdc |
25 Vdc to 1000 Vdc |
| Output setting range |
0 V to 1025 V |
0 V to 1025 V |
0 V to 1025 V |
0 V to 1025 V |
| Output setting resolution |
1 V |
1 V |
1 V |
1 V |
| Output setting accuracy |
±(1.0% of setting + 2 V) |
±(1.0% of setting + 2 V) |
±(1.0% of setting + 2 V) |
±(1.0% of setting + 2 V) |
| Maximum rated load |
1 W (1000 V/1 mA) |
1 W (1000 V/1 mA) |
1 W (1000 V/1 mA) |
1 W (1000 V/1 mA) |
| Maximum rated current |
1 mA |
1 mA |
1 mA |
1 mA |
| Output form |
Floating |
Floating |
Floating |
Floating |
| Isolation voltage |
±500 Vdc, including TOS73 output |
±500 Vdc, including TOS73 output |
±500 Vdc, including TOS73 output |
±500 Vdc, including TOS73 output |
| Ripple at 1 kV, no load |
5 Vp-p or less |
5 Vp-p or less |
5 Vp-p or less |
5 Vp-p or less |
| Ripple at maximum rated load |
10 Vp-p or less |
10 Vp-p or less |
10 Vp-p or less |
10 Vp-p or less |
| Voltage regulation |
1% or less, maximum rated load to no load |
1% or less, maximum rated load to no load |
1% or less, maximum rated load to no load |
1% or less, maximum rated load to no load |
| Constant-current charge current |
2 mA |
2 mA |
2 mA |
2 mA |
| Short-circuit current |
10 mA or less |
10 mA or less |
10 mA or less |
10 mA or less |
| Discharge function |
Constant-current discharge after test; 10 mA typical; 2 µF maximum capacitance within 0.2 s |
Constant-current discharge after test; 10 mA typical; 2 µF maximum capacitance within 0.2 s |
Constant-current discharge after test; 10 mA typical; 2 µF maximum capacitance within 0.2 s |
Constant-current discharge after test; 10 mA typical; 2 µF maximum capacitance within 0.2 s |
| Output voltage rise time with Rise Time off |
1 ms or less |
1 ms or less |
1 ms or less |
1 ms or less |
| Voltage and Resistance Measurement |
| Voltmeter measurement range |
0 Vdc to 1200 Vdc |
0 Vdc to 1200 Vdc |
0 Vdc to 1200 Vdc |
0 Vdc to 1200 Vdc |
| Voltmeter resolution |
0.1 V |
0.1 V |
0.1 V |
0.1 V |
| Voltmeter accuracy |
±(1% of reading + 1 V) |
±(1% of reading + 1 V) |
±(1% of reading + 1 V) |
±(1% of reading + 1 V) |
| Resistance measurement range |
0.001 MΩ to 10.00 GΩ, within 1 mA to 40 nA rated-current range |
0.001 MΩ to 10.00 GΩ, within 1 mA to 40 nA rated-current range |
0.001 MΩ to 10.00 GΩ, within 1 mA to 40 nA rated-current range |
0.001 MΩ to 10.00 GΩ, within 1 mA to 40 nA rated-current range |
| Resistance accuracy, 1 µA < i ≤ 2 mA |
0.001 MΩ to <10.000 MΩ: ±(2% of reading + 5 digits); 10.000 MΩ to <1.000 GΩ: ±2% of reading |
Same |
Same |
Same |
| Resistance accuracy, 200 nA < i ≤ 1000 nA |
10.000 MΩ to <5.000 GΩ: ±5% of reading |
Same |
Same |
Same |
| Resistance accuracy, 100 nA < i ≤ 200 nA |
100.00 MΩ to <5.000 GΩ: ±10% of reading; 5.000 GΩ to <10.000 GΩ: ±20% of reading |
Same |
Same |
Same |
| Resistance accuracy, 40 nA ≤ i ≤ 100 nA |
200.00 MΩ to <10.000 GΩ: ±20% of reading |
Same |
Same |
Same |
| Measurement hold |
Holds the resistance measurement after a test while PASS is displayed |
Same |
Same |
Same |
| Offset cancellation |
Cancels up to 2000 GΩ of unwanted insulation resistance; can be switched off |
Same |
Same |
Same |
| Judgment and Protection |
| Upper-limit setting |
0.02 MΩ to 10.00 GΩ or off |
0.02 MΩ to 10.00 GΩ or off |
0.02 MΩ to 10.00 GΩ or off |
0.02 MΩ to 10.00 GΩ or off |
| Lower-limit setting |
0.01 MΩ to 9.99 GΩ or off |
0.01 MΩ to 9.99 GΩ or off |
0.01 MΩ to 9.99 GΩ or off |
0.01 MΩ to 9.99 GΩ or off |
| Upper-/lower-limit and PASS judgement |
Yes |
Yes |
Yes |
Yes |
| dV judgement |
No |
No |
Yes; Upper-FAIL (dV) after the configured delta voltage is exceeded for the configured consecutive count |
Yes; Upper-FAIL (dV) after the configured delta voltage is exceeded for the configured consecutive count |
| Contact check |
Yes |
Yes |
Yes |
Yes |
| Pre-test short-circuit detection |
Yes |
Yes |
Yes |
Yes |
| Output voltage monitor protection |
Output switches off if voltage exceeds ±(10% of setting + 5 V) |
Same |
Same |
Same |
| Timing and Display |
| Aperture time settings |
16.66 ms, 20 ms, 100 ms |
16.66 ms, 20 ms, 100 ms |
10 ms, 16.66 ms, 20 ms, 100 ms |
10 ms, 16.66 ms, 20 ms, 100 ms |
| Test-time setting range |
0.04 s to 1000.00 s depending on aperture setting; off available |
Same |
0.04 s to 1000.00 s depending on aperture setting; off available |
Same |
| Timing accuracy |
±(100 ppm of setting + 2 ms) below 10 s; ±(100 ppm of setting + 20 ms) at 10 s or above |
Same |
Same |
Same |
| Display |
4.3-inch colour LCD; numeric result display |
4.3-inch colour LCD; numeric result display |
4.3-inch colour LCD; numeric and graph result display |
4.3-inch colour LCD; numeric and graph result display |
| Graphing of test-result data |
No |
No |
Yes |
Yes |
| Data storage / CSV export |
Yes |
Yes |
Yes |
Yes |
| Interfaces |
| RS-232C |
D-sub 9-pin; 9600 to 115200 bps |
Same |
Same |
Same |
| USB device |
USB 2.0 Type B; USBTMC-USB488; 480 Mbps |
Same |
Same |
Same |
| USB host |
USB 2.0 Type A; FAT32 up to 32 GB or exFAT; 480 Mbps |
Same |
Same |
Same |
| LAN |
No |
1000BASE-T/100BASE-TX, Auto-MDIX, IPv4, RJ-45 |
No |
1000BASE-T/100BASE-TX, Auto-MDIX, IPv4, RJ-45 |
| LAN standards and protocols |
Not applicable |
LXI 1.5 Core Specification 2016; HiSLIP and SCPI-RAW |
Not applicable |
LXI 1.5 Core Specification 2016; HiSLIP and SCPI-RAW |
| SIGNAL I/O |
D-sub 25-pin; start/stop, interlock, memory recall, status and judgement monitoring |
Same |
Same |
Same |
| General |
| AC input |
100 Vac to 240 V nominal (90 Vac to 250 V), 47 Hz to 63 Hz |
Same |
Same |
Same |
| Power consumption |
10 VA or less at no load; 50 VA maximum at rated load |
Same |
Same |
Same |
| Operating temperature/humidity |
0°C to 40°C; 20% RH to 80% RH, non-condensing |
Same |
Same |
Same |
| Dimensions |
214 mm W × 80 mm H × 260 mm D (maximum 220 mm × 105 mm × 290 mm) |
Same |
Same |
Same |
| Weight |
Approx. 2.5 kg |
Approx. 2.5 kg |
Approx. 2.5 kg |
Approx. 2.5 kg |
| Supplied accessories |
2.5 m power cord, high-voltage test lead set, Getting Started Guide, Safety Information and RoHS sheet |
Same |
Same |
Same |