The Microtest 6632-50 is a 10Hz to 50MHz Impedance Analyser. The precision impedance analyser 6632 has seven frequency options(10Hz to 1/3/5/10/20/30/50MHz). The equivalent circuit analysis function supports 7 different parameter models (three-element, four-element), it can help customers model the impedance versus frequency characteristics. 6632 Built-in DC Bias voltage ±12V, it can measurement of piezoelectric element admittance circle, and can measure DC bias characteristic of capacitance value. 6632 can use with DC bias current test system. Support RS-232, GPIB, Handler, LAN, USB Host/Device interfaces, PC connection software capabilities to improve the design and test efficiency significantly.

The 6632 supports a variety of accessories that meet customer’s need. Such as dielectric material test fixture that allows PCB designed engineers to reduce the faulty conditions with the comprehensive information of permittivity while designing PCB layouts, liquid dielectric material test fixture for measuring dielectric constant of pure water and air, magnetic material test fixture for measuring of permeability of various toroidal cores or ferrite cores and electromagnetic shielding coating materials……etc.

Features:

  • Signal source frequency range: 10 Hz to 50MHz
  • Basic accuracy up to ±0.08% (typical ±0.05%)
  • ALC function
  • Output impedance 25Ω/100Ω, switchable
  • Support meter mode and list mode, sweep mode, and equivalent circuit analysis(option) function
  • Built-in DC Bias voltage ±12V
  • Measurement of piezoelectric element admittance circle, and can measure DC bias characteristic of capacitance value.
  • Ultra-high measuring speed < 3ms
  • Open circuit/short circuit/load correct function
  • Up to four component parameters can be selected in the electric meter mode. The induction and DCR values can be measured and displayed simultaneously
  • Auto component classification: Comparator function and Handler BIN classification function
  • Can be used with various fixtures, such as: liquid dielectric material test fixture, dielectric material test fixture and magnetic material test fixture….etc.
  • Using with DC bias current test system
  • Support RS-232、GPIB、Handler、LAN、USB Host/Device interfaces
  • PC connection data analysis software is available

Microtest 6632 Impedance Analyser Comparison Table

MODEL Test Frequency Basic Accuracy DC Drive Level Test Time (The Fastest) Output Impedance Measurement Circuit
6632-1 10Hz – 1MHz ± 0.08% (typical ± 0.05%) 1V (fixed) < 3ms 25Ω/100Ω, switchable (Nominal value) Series/Parallel
6632-3 10Hz – 3MHz ± 0.08% (typical ± 0.05%) 1V (fixed) < 3ms 25Ω/100Ω, switchable (Nominal value) Series/Parallel
6632-5 10Hz – 5MHz ± 0.08% (typical ± 0.05%) 1V (fixed) < 3ms 25Ω/100Ω, switchable (Nominal value) Series/Parallel
6632-10 10Hz – 10MHz ± 0.08% (typical ± 0.05%) 1V (fixed) < 3ms 25Ω/100Ω, switchable (Nominal value) Series/Parallel
6632-20 10Hz – 20MHz ± 0.08% (typical ± 0.05%) 1V (fixed) < 3ms 25Ω/100Ω, switchable (Nominal value) Series/Parallel
6632-30 10Hz – 30MHz ± 0.08% (typical ± 0.05%) 1V (fixed) < 3ms 25Ω/100Ω, switchable (Nominal value) Series/Parallel
6632-50 10Hz – 50MHz ± 0.08% (typical ± 0.05%) 1V (fixed) < 3ms 25Ω/100Ω, switchable (Nominal value) Series/Parallel

Description


Documentation

Specification
Test Frequency 6632-1: 10Hz-1MHz
6632-3: 10Hz-3MHz
6632-5: 10Hz-5MHz
6632-10: 10Hz-10MHz
6632-20: 10Hz-20MHz
6632-30: 10Hz-30MHz
6632-50: 10Hz-50MHz
Basic Accuracy ± 0.05%
AC Test Signal Level
Test Signal Voltage Level 10mV  2Vrms(Ro 100Ω)
10mV – 1Vrms(Ro 25Ω)
Voltage Minimum Resolution 1mV
Accuracy ALC ON:6% Voltage ± 2mV
ALC OFF
:10% * Voltage ± 2mV
Test Signal Current Level 200µA – 20mArms
Current Minimum Resolution 10µA
Accuracy ALC ON:6% Current ± 20µA
ALC OFF
:10% * Current ± 20µA
DC Test Signal Level 1V (Fixed)
Test Time (The fastest) < 3mS
Output Impedance 25Ω/100Ω, switchable (Nominal value)
Parameters |Z|、|Y|、θ、X、R、G、B、L、D、Q、DCR、C、 ESR、ε and μr
Bias Can be using with DC Bias
Measurement Mode Meter mode, list mode, sweep mode, and optional equivalent circuit analysis function(S model)
Correction Open Circuit/ Short Circuit/Load Correction
Measurement Circuit Series/Parallel
List mode 50 groups of list mode setting (Each group contains up to 15 steps)
Equivalent Circuit Analysis Three elements, Four elements (option)
DC bias voltage -12 to +12V, 0.3% ±1.5mV, 100Hz to 30MHz
BIN 9 Level
Comparator ABS、ΔABS、Δ%、OFF
Internal Storage 100 sets LCR setting documents, 50 groups of list mode setting
USB Storage LCR setting documents, list mode setting document, BMP graphics, Sweep screen and test result data
Interface RS-232、GPIB、Handler、LAN、USB Host/Device
Option PC connection software
Equivalent Circuit Analysis
Power Supply Voltage 90V ~ 264Vac
Frequency 47Hz ~ 63Hz
Power Consumption Low power consumption: Maximum 30W (Nominal value)
Display 7.0”TFT, 800×480 colour screen
Environment Temperature: 10℃ ~ 40℃   Humidity: ≦90%
Dimension (W*H*D) 336x147x340 mm
Weight 3.95kg
Parameters Measurement Ranges
|Z| 0.000 mΩ – 9999.99 MΩ
R, X ± 0.000 mΩ – 9999.99 MΩ
|Y| 0.00000 uS – 999.999 kS
G, B ± 0.00000 uS – 999.999 kS
θRAD ± 0.00000 – 3.14159
θDEG ± 0.000° – 180.000°
Cs, Cp ± 0.00000 pF – 9999.99 F
Ls, Lp ± 0.00 nH – 9999.99 kH
D 0.00000 – 9999.99
Q 0.00 – 9999.99
Δ ± 0.00% – 9999.99%
Rdc 0.00 mΩ – 99.9999 MΩ
εr’ εr” 0 – 10000
μr’ μr'” 0 – 10000
Applications
Passive Components Capacitor, Inductor, Resistor, Transformer, Ceramic resonator, Quartz Crystal
Semiconductor Components The CV characteristics analysis of varactor diodes, Diodes
Dielectric Material Estimation on permittivity and consumption tangent of plastic, ceramic and PCB
Other Components Estimation of the impedance of PCB components

Note*

Specifications are liable to change. For the full list of Specifications, please see the products datasheet located under the ‘Documentation’ tab.

Mandatory Accessories …

  • UK Power Cord

    .
    Read more